Mettler Toledo Balance

by K-Dense-AIversion 0.1.3i★ 32 (Stars for the whole repo)iRuns on your computerRepository ↗
Registry nameio.github.K-Dense-AI/labmcp-mettler-toledo

How to install

What it does

MCP server for Mettler Toledo laboratory balances (MT-SICS): weigh, tare, zero, log drift, adjust.

  • Connects Runs on your computer
  • Account or key No key needed
  • Category Other

Why it ranks here

Lists are ordered by a usage score (out of 100) built from these parts

Popularity11/80GitHub stars or weekly downloads: the larger counts 60, the smaller 20
Freshness7/12Full marks for a release or code change within 30 days
Still answering0/8The cloud endpoint answered at the last check
Usage score19/100

Quality: Passed 3/3 basic checks (shown, not used for ordering) · Momentum: No data yet (Used only on the Trending tab)

Install

Pick the AI app you use, then copy the settings.

Let your AI install it with one sentence

Please install Mettler Toledo Balance. Follow https://mcp.popnx.com/i/k-dense-ai-labmcp-mettler-toledo.md, show me the commands and wait for my OK before running them.

Paste it into an AI that can work on your computer (such as Claude Code or Cursor). It reads our steps, shows you the commands and waits for your OK. No key, no sign-up. · See the full text your AI reads

Passed 3/3 checks · Runs on your computer · ✓ Commands checked Nov 14, 2026

First: which AI do you use?

Pick your app and the steps appear below.

Works with Claude, with some setup first

Install Python (uv) on your computer first (one time) · Only the Claude desktop app (Windows or Mac), not the web or phone

Let AI install it

Copy this message into a Claude chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Mettler Toledo Balance for Claude on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "mcpServers": { "labmcp-mettler-toledo": { "command": "uvx", "args": [ "labmcp-mettler-toledo" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Mettler Toledo Balance tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Open the Claude desktop app’s settings file

    SettingsDeveloperEdit Config

  3. Paste these settings and save the file

    {
      "mcpServers": {
        "labmcp-mettler-toledo": {
          "command": "uvx",
          "args": [
            "labmcp-mettler-toledo"
          ]
        }
      }
    }
  4. Quit Claude and open it again

  5. Try it

    Use Mettler Toledo Balance and tell me what you can do for me, then show me one thing

    ✓ If Claude says it is using Mettler Toledo Balance or shows a Mettler Toledo Balance tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Anthropic’s help page on Nov 11, 2026 · your screen may differ slightly by app version

ChatGPT cannot connect Mettler Toledo Balance yet

Mettler Toledo Balance runs on your computer, but ChatGPT can only connect servers that run on the web

What you can do instead

A · Use an app that can connect Mettler Toledo Balance

  • Install Python (uv) on your computer first (one time) · Only the Claude desktop app (Windows or Mac), not the web or phone
  • Install Python (uv) on your computer first (one time)
  • Install Python (uv) on your computer first (one time)
  • Install Python (uv) on your computer first (one time)

B · Others in Other that work with ChatGPT

Picked from the most used in the same category, only ones that connect by address with no key

Works with Claude Code, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a Claude Code chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Mettler Toledo Balance for Claude Code on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): claude mcp add --scope user labmcp-mettler-toledo -- uvx labmcp-mettler-toledo 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Mettler Toledo Balance tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Open Terminal

    Mac: press Command+Space and type Terminal · Windows: open PowerShell

  3. Paste this command and press Enter

    claude mcp add --scope user labmcp-mettler-toledo -- uvx labmcp-mettler-toledo
  4. Type /mcp in Claude Code

    You should see Mettler Toledo Balance listed as connected

  5. Try it

    Use Mettler Toledo Balance and tell me what you can do for me, then show me one thing

    ✓ If Claude Code says it is using Mettler Toledo Balance or shows a Mettler Toledo Balance tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Anthropic’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Works with Cursor, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a Cursor chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Mettler Toledo Balance for Cursor on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "mcpServers": { "labmcp-mettler-toledo": { "command": "uvx", "args": [ "labmcp-mettler-toledo" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Mettler Toledo Balance tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Click this button to open Cursor

    Open in Cursor

    Cursor asks you to confirm; click Install

  3. Open the AI chat in Cursor in Agent mode

  4. Try it

    Use Mettler Toledo Balance and tell me what you can do for me, then show me one thing

    ✓ If Cursor says it is using Mettler Toledo Balance or shows a Mettler Toledo Balance tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Cursor’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Works with VS Code, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a VS Code chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Mettler Toledo Balance for VS Code on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "servers": { "labmcp-mettler-toledo": { "type": "stdio", "command": "uvx", "args": [ "labmcp-mettler-toledo" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Mettler Toledo Balance tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Click this button to open VS Code

    Open in VS Code

    VS Code asks you to confirm; click Install

  3. Open the AI chat in VS Code in Agent mode

  4. Try it

    Use Mettler Toledo Balance and tell me what you can do for me, then show me one thing

    ✓ If VS Code says it is using Mettler Toledo Balance or shows a Mettler Toledo Balance tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against VS Code’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Other apps that use MCP, such as Windsurf, Zed or Gemini CLI, each add servers differently. See how in the app list.

See every app that uses MCP →

JSON settings are under “For developers” below

Stuck? Ask people who use Mettler Toledo Balance💬 Discussion →
For developers: Terminal commands and JSON settings

Installed on your computer

Claude Code

claude mcp add --scope user labmcp-mettler-toledo -- uvx labmcp-mettler-toledo

Claude Desktop (claude_desktop_config.json)

{
  "mcpServers": {
    "labmcp-mettler-toledo": {
      "command": "uvx",
      "args": [
        "labmcp-mettler-toledo"
      ]
    }
  }
}

Cursor (~/.cursor/mcp.json)

{
  "mcpServers": {
    "labmcp-mettler-toledo": {
      "command": "uvx",
      "args": [
        "labmcp-mettler-toledo"
      ]
    }
  }
}

VS Code (mcp.json)

{
  "servers": {
    "labmcp-mettler-toledo": {
      "type": "stdio",
      "command": "uvx",
      "args": [
        "labmcp-mettler-toledo"
      ]
    }
  }
}

Values you fill in

  • LABMCP_ADDRESS env
    Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE env
    Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY env
    Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS env
    Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

These settings are generated from the maker’s registry entry. If the README says otherwise, follow the README.

At a glance

GitHub starsi
32 (Stars for the whole repo)
This repo holds several MCP listings; its stars count once in rankings
Downloads per weeki
44/wk
Licencei
Apache-2.0
Connection
Runs on your computer
Latest versioni
0.1.3
Categoryi
Other
Published to registryi
Sep 27, 2026
Code last changedi
Sep 27, 2026
Maker
K-Dense-AI
Packages
pypi: labmcp-mettler-toledo
Website
github.com
Registry status
Active

Which apps can use it

Based on how this server connects and on each app’s own docs, checked Nov 10, 2026.

Run on your computer in

ClaudeGooseLM StudioMsty StudioCherry StudioCursorVS Code + GitHub CopilotVisual StudioWindsurf (Devin Desktop)ZedKiroClineContinueJetBrains AI AssistantClaude CodeCodex CLIGemini CLIOpenCodenodecodeWarpLibreChatRaycast

All apps, with pros and cons →

Related guides

What it can do

Runs on your computer; the tool list appears in your AI app after you install it.

Basic checksi

We check information anyone can see; we have not read the code. Checks that do not apply are left out of the count. · Checked Nov 20, 2026

  • ✓ Pass

    Maker matches the repo

    The registry name was verified with the GitHub account K-Dense-AI, which owns the repo

  • ✓ Pass

    Still maintained

    Code last changed 2 months ago

  • ✓ Pass

    Has a licence

    Uses Apache-2.0

  • – Not applicable

    Endpoint responds

    This one runs on your computer; there is no internet endpoint

  • – Not applicable

    Sign-in with OAuth

    Runs on your computer and uses local settings instead

  • … Not checked yet

    No hidden instructions in tool descriptions

    Runs on your computer; we can only read tool lists from internet endpoints

Published vulnerabilities (OSV.dev)

No entries on OSV.dev for labmcp-mettler-toledo@0.1.3

Checks only this package version, not every dependency · checked Nov 16, 2026

What we check → · JSON

Makers: ask for a re-check or report wrong info
For makers: add a badge to your README

If you made this, paste this code into your README. The badge links back to this page.

[![Featured on PopNx MCP](https://mcp.popnx.com/badge/k-dense-ai-labmcp-mettler-toledo.svg)](https://mcp.popnx.com/en/servers/k-dense-ai-labmcp-mettler-toledo)

Discussion

Ask, share a tip or report that it does not work. Sorted by what people marked helpful.

No posts yet. Be the first to ask or share.

Posts here are written by users, not the PopNx team.

Sign in to post

User reviews

No reviews yet

Sign in to write a review

Compared with others in Otheri

  • More stars than 96% of the other 20815 servers in this category
  • Published to the registry more recently than 45% of the other 20815 servers in this category
  • Basic checks equal to or better than 99% of the other 20815 servers in this category

Compared with every active server in this category in the registry. Stars count only when the publisher owns the repo.

Common questions

Is Mettler Toledo Balance safe?

It passed 3 of 3 applicable basic checks. We have not read the code. Before adding a key or token, see what access the server asks for and use a key with only the permissions you need.

What keys or codes does it need?

The registry does not list a required key, though some servers ask you to sign in on first use.

Does it run on my computer or in the cloud?

It runs on your computer. Your AI app starts it when needed, and data does not pass through the maker’s servers (unless it calls an online service itself).

Is it still maintained?

The GitHub repo last changed 2 months ago, and the latest version (0.1.3) reached the registry 2 months ago.

Which apps can use it?

Apps that support MCP, such as Claude Desktop, Claude Code, Cursor and VS Code. You also need Python (uv) on your computer.

What can I use instead?

In the same category: World Monitor, openstatus, Oh My Posh Validator, Strata, Jitsu

Similar alternatives

Same category, other makers, commands already checked. Official ones first, then by stars.

Reading a card:↓ Downloads per weeki★ GitHub starsiThree pillars: reviews · checks · usageiUpdated = latest version published to the registryi

World Monitor

Real-time global intelligence on conflict, markets, energy, trade, shipping and cyber. 100 tools.

by koala73Widely used★ 90.9KUpdated 4 days ago

Runs in the cloud✓ service owner · worldmonitor.app

openstatus

Manage monitors, status pages, incidents and maintenances in your openstatus workspace.

by openstatusHQWidely used★ 9.3KUpdated 2 months ago

Runs in the cloud✓ service owner · openstatus.dev

Oh My Posh Validator

Validate oh-my-posh configurations and segment snippets against the official schema.

by JanDeDobbeleerWidely used★ 23.8K Stars for the whole repoUpdated 31 days ago

Runs in the cloud✓ service owner · ohmyposh.dev

Strata

MCP server for progressive tool usage at any scale (see https://klavis.ai)

by Klavis-AIGrowing★ 5.8KUpdated 14 months ago

Runs in the cloudRepo archived✓ klavis.ai

Jitsu

Manage Jitsu data pipelines: destinations, streams, connections, functions, live events.

by jitsucomWidely used★ 5.1KUpdated 4 months ago

Runs in the cloud✓ service owner · jitsu.com
Compared with similar ones
This one
Mettler Toledo Balance
World MonitoropenstatusOh My Posh ValidatorStrataJitsu
Runs whereInstalls on your computerRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to install
Needs a keyNoNoNoNoNoNo
Add with + in ClaudeNoNoNoNoNoNo
Basic checks3/34/44/44/41/44/4
LicenseApache-2.0AGPL-3.0AGPL-3.0MITApache-2.0MIT
PriceNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registry

README from the maker

Open the README

Copied from the maker’s repo as written. PopNx MCP has not reviewed or endorsed it. HTML, images and scripts are removed. Links lead off this site.

Mettler Toledo Balance — MCP Server

Let an AI agent weigh samples, tare containers, log drift and evaporation, run internal adjustments and drive the draft shield on Mettler Toledo balances through the open MT-SICS interface command set.

Packagelabmcp-mettler-toledo
InstrumentsExcellence XPR/XSR, XP/XS, XA/XE, MS/ML, NewClassic ME/MS, and other MT-SICS balances and weighing terminals
InterfacesRS-232, USB (virtual COM port), Ethernet (XPR/XSR, or any serial-to-Ethernet adapter)
ProtocolMT-SICS levels 0–2 (reference manual)
Status🧪 simulated: tested against a wire-level MT-SICS simulator, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-mettler-toledo --simulate --check

Connect your balance

  1. Balance setup: in the balance menu, set the interface (RS232/USB/Ethernet) to MT-SICS (sometimes called "Host" or "Dialog"). Factory settings are 9600 baud, 8 data bits, no parity, 1 stop bit, Xon/Xoff.
  2. Find the port: uvx labmcp ports
  3. Test the connection:
   uvx labmcp-mettler-toledo --address /dev/ttyUSB0 --check          # Linux

uvx labmcp-mettler-toledo --address /dev/tty.usbserial-XXXX --check # macOS uvx labmcp-mettler-toledo --address COM4 --check # Windows uvx labmcp-mettler-toledo --address tcp://192.168.1.60:8001 --check # Ethernet

   Non-default settings go in the address: `serial://COM4?baudrate=19200&xonxoff=false`.

Add to your MCP client

Claude Code

claude mcp add balance -- uvx labmcp-mettler-toledo --address /dev/ttyUSB0

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)

{

"mcpServers": { "balance": { "command": "uvx", "args": ["labmcp-mettler-toledo", "--address", "/dev/ttyUSB0"] } } }

Add --read-only to allow weighing but block taring, zeroing and door movement. For other clients, generate the snippet with uvx labmcp config mettler-toledo --address /dev/ttyUSB0 --client vscode (also cursor, codex, claude-code).

Tools

ToolKindDescription
clear_tare🎛 controlClear the tare memory (tare = 0).
get_command_log👁 readReturn the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info👁 readReport which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_draft_shield👁 readReport the position of the motorised draft-shield doors (Excellence/XPR balances only).
get_tare👁 readReturn the weight currently stored in the tare memory.
log_weight_series👁 readRecord a series of immediate (unfiltered) readings to monitor drift, evaporation, moisture uptake or stabilisation. Returns every reading plus summary statistics.
read_temperature👁 readRead the balance's internal temperature probe(s) in °C (MT-SICS M28, if supported).
read_weight👁 readRead the current net weight from the balance.
reconnect🛑 safetyClose and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
reset_balance🛑 safetyAbort whatever the balance is doing (an adjustment, a repeating weight stream, a pending command) and reset it to its power-on state without zeroing (MT-SICS @). Clears the tare. Motorised draft-shield doors have their own obstruction detection and reverse if blocked.
run_internal_adjustment🎛 controlAdjust (calibrate) the balance with its built-in reference weight (MT-SICS C3). The pan must be empty and the balance undisturbed. Takes roughly 1-3 minutes.
set_draft_shield⚠️ hazardOpen or close the motorised draft-shield doors. Make sure nothing (and no one's fingers) is in the way of the doors.
set_tare_preset🎛 controlPreset a known tare weight (e.g. a container weighed earlier).
show_message🎛 controlShow a short message on the balance display (e.g. 'Add sample 3'). Use show_weight_display to return to the normal weight display.
show_weight_display🎛 controlSwitch the balance display back to showing the weight.
tare🎛 controlTare the balance: store the current load (e.g. an empty container) as the tare weight. Returns the stored tare weight.
zero🎛 controlZero the balance with the current load. Also clears the tare memory.

get_connection_info, get_command_log and reconnect are built into every LabMCP server.

Safety limits

LimitDefaultMeaning
max_series_duration_s600 sLongest weight-logging series an agent may start

Override at launch: --limit max_series_duration_s=3600. Whatever the limit says, a single series is capped at one hour (the tool's time limit).

Example prompts

  • "Tare the balance, then tell me when the reading is stable after I add the sample."
  • "Log the weight every 10 s for 5 minutes and tell me the evaporation rate in mg/min."
  • "Weigh out ~250 mg: tell me how much more to add, reading every few seconds, until I'm within ±2 mg."
  • "Run an internal adjustment and confirm it succeeded before we start the calibration curve."

Notes

  • read_weight(stable=True) sends S, which waits for stability. If the balance cannot settle (draughts, vibration), it returns S I. The server reports this as an error, so retry or use stable=false.
  • Weights are in the balance's unit 1 (normally grams). Units shown on the display (unit 2) don't affect the interface.
  • Draft-shield (WS) and temperature (M28) commands exist only on balances with that hardware; other balances answer ES/L, which the server reports clearly.
  • run_internal_adjustment waits for the final C3 A without blocking the connection: reset_balance aborts a running adjustment, and other commands are refused until it ends (their replies would be mixed up with the result).
  • Before each command the server discards anything the balance sent unasked (e.g. a print-key transmission) and checks that the reply belongs to the command, so one late reply cannot shift every later reply by one.
  • show_message accepts printable ASCII only, and set_tare_preset only a unit symbol: a line break would otherwise reach the balance as a second MT-SICS command.

Hardware verification

ModelFirmwareInterfaceVerified byDate
*none yet: be the first*