Prototype · Preview: every server in the official registry, synced every 30 minutes. Sign in with Google to write reviews.Prototype · updated every 30 min
PopNx MCPMCP servers for people who use AI · explained in plain words

Thorlabs Optical Power Meter

by K-Dense-AIversion 0.1.2i★ 32 (Stars for the whole repo)iRuns on your computerRepository ↗
Registry nameio.github.K-Dense-AI/labmcp-thorlabs-pm

How to install

What it does

MCP server for Thorlabs optical power meters (PM100 family, PM400, PM101, PM5020) over SCPI.

  • Connects Runs on your computer
  • Account or key No key needed
  • Category Other

Why it ranks here

Lists are ordered by a usage score (out of 100) built from these parts

Popularity0/80GitHub stars or weekly downloads: the larger counts 60, the smaller 20
Freshness7/12Full marks for a release or code change within 30 days
Still answering0/8The cloud endpoint answered at the last check
Usage score7/100

Quality: Passed 3/3 basic checks (shown, not used for ordering) · Momentum: No data yet (Used only on the Trending tab)

Install

Pick the AI app you use, then copy the settings.

Let your AI install it with one sentence

Please install Thorlabs Optical Power Meter. Follow https://mcp.popnx.com/i/k-dense-ai-labmcp-thorlabs-pm.md, show me the commands and wait for my OK before running them.

Paste it into an AI that can work on your computer (such as Claude Code or Cursor). It reads our steps, shows you the commands and waits for your OK. No key, no sign-up. · See the full text your AI reads

Passed 3/3 checks · Runs on your computer · ✓ Commands checked Nov 14, 2026

First: which AI do you use?

Pick your app and the steps appear below.

Works with Claude, with some setup first

Install Python (uv) on your computer first (one time) · Only the Claude desktop app (Windows or Mac), not the web or phone

Let AI install it

Copy this message into a Claude chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Thorlabs Optical Power Meter for Claude on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "mcpServers": { "labmcp-thorlabs-pm": { "command": "uvx", "args": [ "labmcp-thorlabs-pm" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Thorlabs Optical Power Meter tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Open the Claude desktop app’s settings file

    SettingsDeveloperEdit Config

  3. Paste these settings and save the file

    {
      "mcpServers": {
        "labmcp-thorlabs-pm": {
          "command": "uvx",
          "args": [
            "labmcp-thorlabs-pm"
          ]
        }
      }
    }
  4. Quit Claude and open it again

  5. Try it

    Use Thorlabs Optical Power Meter and tell me what you can do for me, then show me one thing

    ✓ If Claude says it is using Thorlabs Optical Power Meter or shows a Thorlabs Optical Power Meter tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Anthropic’s help page on Nov 11, 2026 · your screen may differ slightly by app version

ChatGPT cannot connect Thorlabs Optical Power Meter yet

Thorlabs Optical Power Meter runs on your computer, but ChatGPT can only connect servers that run on the web

What you can do instead

A · Use an app that can connect Thorlabs Optical Power Meter

  • Install Python (uv) on your computer first (one time) · Only the Claude desktop app (Windows or Mac), not the web or phone
  • Install Python (uv) on your computer first (one time)
  • Install Python (uv) on your computer first (one time)
  • Install Python (uv) on your computer first (one time)

B · Others in Other that work with ChatGPT

Picked from the most used in the same category, only ones that connect by address with no key

Works with Claude Code, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a Claude Code chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Thorlabs Optical Power Meter for Claude Code on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): claude mcp add --scope user labmcp-thorlabs-pm -- uvx labmcp-thorlabs-pm 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Thorlabs Optical Power Meter tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Open Terminal

    Mac: press Command+Space and type Terminal · Windows: open PowerShell

  3. Paste this command and press Enter

    claude mcp add --scope user labmcp-thorlabs-pm -- uvx labmcp-thorlabs-pm
  4. Type /mcp in Claude Code

    You should see Thorlabs Optical Power Meter listed as connected

  5. Try it

    Use Thorlabs Optical Power Meter and tell me what you can do for me, then show me one thing

    ✓ If Claude Code says it is using Thorlabs Optical Power Meter or shows a Thorlabs Optical Power Meter tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Anthropic’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Works with Cursor, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a Cursor chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Thorlabs Optical Power Meter for Cursor on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "mcpServers": { "labmcp-thorlabs-pm": { "command": "uvx", "args": [ "labmcp-thorlabs-pm" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Thorlabs Optical Power Meter tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Click this button to open Cursor

    Open in Cursor

    Cursor asks you to confirm; click Install

  3. Open the AI chat in Cursor in Agent mode

  4. Try it

    Use Thorlabs Optical Power Meter and tell me what you can do for me, then show me one thing

    ✓ If Cursor says it is using Thorlabs Optical Power Meter or shows a Thorlabs Optical Power Meter tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against Cursor’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Works with VS Code, with some setup first

Install Python (uv) on your computer first (one time)

Let AI install it

Copy this message into a VS Code chat. The AI checks your computer, tells you which file it will change, and waits for your OK every time.

Please install the MCP server Thorlabs Optical Power Meter for VS Code on this computer, in this order: 1) Check uv with uv --version. If it is missing, tell me to install it from https://docs.astral.sh/uv/getting-started/installation/; do not install it yourself. 2) Use only this config (already checked by PopNx MCP): { "servers": { "labmcp-thorlabs-pm": { "type": "stdio", "command": "uvx", "args": [ "labmcp-thorlabs-pm" ] } } } 3) Tell me the full path of the file you will change, or the command you will run, and wait for my OK. 4) No key is needed. 5) Do what we agreed, then check that the Thorlabs Optical Power Meter tools really show up.

This message has no key in it. If a key is needed, the AI will ask you to paste it yourself into the file or the app’s field.

  1. Install uv (if you don’t have it)

    Already installed? Check first · Open Terminal, type this command and press Enter

    uv --version
    • A version number such as uv … = installed, skip this step
    • "command not found" (on Windows: "is not recognized") = not installed yet, or installed but Terminal needs to be closed and reopened
    How to open Terminal

    Mac: press ⌘ + Space, type Terminal, press Enter

    Windows: open Start, type Windows PowerShell, then Open

    Not installed: uv has no click-through installer. Copy the one-line command for your system (macOS or Windows) from uv’s official page into Terminal · no separate Python install needed; uv downloads it when required

    Open the uv download page ↗

    When it’s done, close Terminal, open it again and run the same command. You should see a version number: uv --version

    Install once; it works for everything that needs uv

  2. Click this button to open VS Code

    Open in VS Code

    VS Code asks you to confirm; click Install

  3. Open the AI chat in VS Code in Agent mode

  4. Try it

    Use Thorlabs Optical Power Meter and tell me what you can do for me, then show me one thing

    ✓ If VS Code says it is using Thorlabs Optical Power Meter or shows a Thorlabs Optical Power Meter tool in the answer, you are connected

4 optional setting(s) (works without them)

The command above works without these. Add them only if you want to change something, following the maker’s description.

  • LABMCP_ADDRESS Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

Menu names checked against VS Code’s help page on Nov 11, 2026 · your screen may differ slightly by app version

Other apps that use MCP, such as Windsurf, Zed or Gemini CLI, each add servers differently. See how in the app list.

See every app that uses MCP →

JSON settings are under “For developers” below

Stuck? Ask people who use Thorlabs Optical Power Meter💬 Discussion →
For developers: Terminal commands and JSON settings

Installed on your computer

Claude Code

claude mcp add --scope user labmcp-thorlabs-pm -- uvx labmcp-thorlabs-pm

Claude Desktop (claude_desktop_config.json)

{
  "mcpServers": {
    "labmcp-thorlabs-pm": {
      "command": "uvx",
      "args": [
        "labmcp-thorlabs-pm"
      ]
    }
  }
}

Cursor (~/.cursor/mcp.json)

{
  "mcpServers": {
    "labmcp-thorlabs-pm": {
      "command": "uvx",
      "args": [
        "labmcp-thorlabs-pm"
      ]
    }
  }
}

VS Code (mcp.json)

{
  "servers": {
    "labmcp-thorlabs-pm": {
      "type": "stdio",
      "command": "uvx",
      "args": [
        "labmcp-thorlabs-pm"
      ]
    }
  }
}

Values you fill in

  • LABMCP_ADDRESS env
    Instrument address, e.g. serial:///dev/ttyUSB0 or tcp://192.168.1.50:5025
  • LABMCP_SIMULATE env
    Set to 1 to use the built-in simulator (no hardware)
  • LABMCP_READ_ONLY env
    Set to 1 to disable all state-changing tools
  • LABMCP_LIMITS env
    Safety limit overrides, e.g. max_temperature_c=80

Descriptions are as written by the maker.

These settings are generated from the maker’s registry entry. If the README says otherwise, follow the README.

At a glance

GitHub starsi
32 (Stars for the whole repo)
This repo holds several MCP listings; its stars count once in rankings
Downloads per weeki
34/wk
Licencei
Apache-2.0
Connection
Runs on your computer
Latest versioni
0.1.2
Categoryi
Other
Published to registryi
Sep 27, 2026
Code last changedi
Sep 27, 2026
Maker
K-Dense-AI
Packages
pypi: labmcp-thorlabs-pm
Website
github.com
Registry status
Active

Which apps can use it

Based on how this server connects and on each app’s own docs, checked Nov 10, 2026.

Run on your computer in

ClaudeGooseLM StudioMsty StudioCherry StudioCursorVS Code + GitHub CopilotVisual StudioWindsurf (Devin Desktop)ZedKiroClineContinueJetBrains AI AssistantClaude CodeCodex CLIGemini CLIOpenCodenodecodeWarpLibreChatRaycast

All apps, with pros and cons →

Related guides

What it can do

Runs on your computer; the tool list appears in your AI app after you install it.

Basic checksi

We check information anyone can see; we have not read the code. Checks that do not apply are left out of the count. · Checked Nov 19, 2026

  • ✓ Pass

    Maker matches the repo

    The registry name was verified with the GitHub account K-Dense-AI, which owns the repo

  • ✓ Pass

    Still maintained

    Code last changed 2 months ago

  • ✓ Pass

    Has a licence

    Uses Apache-2.0

  • – Not applicable

    Endpoint responds

    This one runs on your computer; there is no internet endpoint

  • – Not applicable

    Sign-in with OAuth

    Runs on your computer and uses local settings instead

  • … Not checked yet

    No hidden instructions in tool descriptions

    Runs on your computer; we can only read tool lists from internet endpoints

Published vulnerabilities (OSV.dev)

No entries on OSV.dev for labmcp-thorlabs-pm@0.1.2

Checks only this package version, not every dependency · checked Nov 16, 2026

What we check → · JSON

Makers: ask for a re-check or report wrong info
For makers: add a badge to your README

If you made this, paste this code into your README. The badge links back to this page.

[![Featured on PopNx MCP](https://mcp.popnx.com/badge/k-dense-ai-labmcp-thorlabs-pm.svg)](https://mcp.popnx.com/en/servers/k-dense-ai-labmcp-thorlabs-pm)

Discussion

Ask, share a tip or report that it does not work. Sorted by what people marked helpful.

No posts yet. Be the first to ask or share.

Posts here are written by users, not the PopNx team.

Sign in to post

User reviews

No reviews yet

Sign in to write a review

Compared with others in Otheri

  • More stars than 0% of the other 20653 servers in this category
  • Published to the registry more recently than 45% of the other 20653 servers in this category
  • Basic checks equal to or better than 99% of the other 20653 servers in this category

Compared with every active server in this category in the registry. Stars count only when the publisher owns the repo.

Common questions

Is Thorlabs Optical Power Meter safe?

It passed 3 of 3 applicable basic checks. We have not read the code. Before adding a key or token, see what access the server asks for and use a key with only the permissions you need.

What keys or codes does it need?

The registry does not list a required key, though some servers ask you to sign in on first use.

Does it run on my computer or in the cloud?

It runs on your computer. Your AI app starts it when needed, and data does not pass through the maker’s servers (unless it calls an online service itself).

Is it still maintained?

The GitHub repo last changed 2 months ago, and the latest version (0.1.2) reached the registry 2 months ago.

Which apps can use it?

Apps that support MCP, such as Claude Desktop, Claude Code, Cursor and VS Code. You also need Python (uv) on your computer.

What can I use instead?

In the same category: World Monitor, openstatus, Oh My Posh Validator, Strata, Jitsu

Similar alternatives

Same category, other makers, commands already checked. Official ones first, then by stars.

Reading a card:↓ Downloads per weeki★ GitHub starsiThree pillars: reviews · checks · usageiUpdated = latest version published to the registryi

World Monitor

Real-time global intelligence on conflict, markets, energy, trade, shipping and cyber. 100 tools.

by koala73Widely used★ 90.9KUpdated 3 days ago

Runs in the cloud✓ service owner · worldmonitor.app

openstatus

Manage monitors, status pages, incidents and maintenances in your openstatus workspace.

by openstatusHQWidely used★ 9.3KUpdated 2 months ago

Runs in the cloud✓ service owner · openstatus.dev

Oh My Posh Validator

Validate oh-my-posh configurations and segment snippets against the official schema.

by JanDeDobbeleerWidely used★ 23.8K Stars for the whole repoUpdated 30 days ago

Runs in the cloud✓ service owner · ohmyposh.dev

Strata

MCP server for progressive tool usage at any scale (see https://klavis.ai)

by Klavis-AIGrowing★ 5.8KUpdated 14 months ago

Runs in the cloudRepo archived✓ klavis.ai

Jitsu

Manage Jitsu data pipelines: destinations, streams, connections, functions, live events.

by jitsucomWidely used★ 5.1KUpdated 4 months ago

Runs in the cloud✓ service owner · jitsu.com
Compared with similar ones
This one
Thorlabs Optical Power Meter
World MonitoropenstatusOh My Posh ValidatorStrataJitsu
Runs whereInstalls on your computerRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to installRuns in the cloud, nothing to install
Needs a keyNoNoNoNoNoNo
Add with + in ClaudeNoNoNoNoNoNo
Basic checks3/34/44/44/41/44/4
LicenseApache-2.0AGPL-3.0AGPL-3.0MITApache-2.0MIT
PriceNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registryNo cost listed in the registry

README from the maker

Open the README

Copied from the maker’s repo as written. PopNx MCP has not reviewed or endorsed it. HTML, images and scripts are removed. Links lead off this site.

Thorlabs Optical Power Meter — MCP Server

Let an AI agent read laser power, log stability and warm-up drift, set the wavelength correction, averaging and range, and dark-zero the sensor on Thorlabs optical power meters through their documented SCPI command set over USB (USBTMC / VISA).

Packagelabmcp-thorlabs-pm
InstrumentsPM100D, PM100A, PM100USB, PM400 consoles with C-series photodiode and thermal sensors. PM101 and PM5020 use the same documented commands but are unverified (see Notes)
InterfacesUSB (USBTMC, via VISA)
ProtocolSCPI: PM100D Operation Manual 17654-D02 §6.4, PM100USB Operation Manual 19570-D02 §6.3, PM400 Operating Manual v1.1 §5.5, and Thorlabs' per-model SCPI command references
Status🧪 simulated: tested against a wire-level SCPI simulator, not yet verified on hardware. Report a hardware test

Try it without hardware

uvx labmcp-thorlabs-pm --simulate --check

uvx labmcp-thorlabs-pm --simulate --option sim_sensor=S302C --check # thermal sensor head

The simulator is a PM100D with an S120C photodiode (or S302C thermopile) looking at a 1.234 mW, 632.8 nm HeNe laser. Setting the wrong wavelength on the photodiode gives a proportionally wrong reading, as on the real bench.

Connect your power meter

  1. Driver. Thorlabs meters enumerate as USB Test & Measurement Class (USBTMC) devices, vendor ID 0x1313.
  • Windows: the Thorlabs driver (TLPM / Thorlabs VISA) is *not* visible to PyVISA. Install the free NI-VISA runtime, then switch the meter to the NI-VISA driver with the Thorlabs Driver Switcher (installed with the Thorlabs Optical Power Monitor software), and start the server with ?backend=@ivi (see below). This is the route Thorlabs documents for PyVISA.
  • Linux: pyvisa-py talks USBTMC directly (installed with this package). Give your user access to the device, e.g. /etc/udev/rules.d/99-thorlabs-pm.rules: SUBSYSTEM=="usb", ATTR{idVendor}=="1313", MODE="0666" then sudo udevadm control --reload-rules and re-plug the meter. NI-VISA also works (?backend=@ivi).
  • macOS: pyvisa-py + libusb (brew install libusb); NI-VISA is an alternative.
  1. Find the resource string: uvx labmcp ports (or python -m pyvisa info). It looks like USB0::0x1313::0x8078::P0012345::INSTR (the product ID depends on the model: PM100D 0x8078, PM100USB 0x8072).
  2. Test the connection:
   uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR --check

uvx labmcp-thorlabs-pm --address "visa://USB0::0x1313::0x8078::P0012345::INSTR?backend=@ivi" --check # NI-VISA

   For the second input of a PM5020 add `--option channel=2`.

Add to your MCP client

Claude Code

claude mcp add power-meter -- uvx labmcp-thorlabs-pm --address USB0::0x1313::0x8078::P0012345::INSTR

Claude Desktop / Cursor / Windsurf (claude_desktop_config.json, .cursor/mcp.json, …)

{

"mcpServers": { "power-meter": { "command": "uvx", "args": ["labmcp-thorlabs-pm", "--address", "USB0::0x1313::0x8078::P0012345::INSTR"] } } }

Add --read-only to allow readings but block wavelength, averaging, range and zero changes. For other clients, generate the snippet with uvx labmcp config thorlabs-pm --address --client vscode (also cursor, codex, claude-code).

Tools

ToolKindDescription
get_command_log👁 readReturn the most recent raw commands sent to / replies received from the instrument (newest last). Useful for debugging and for recording what was done.
get_connection_info👁 readReport which instrument is connected (identity, address, simulated or real), whether the server is read-only, and the active safety limits. Call this first.
get_device_info👁 readIdentify the meter console and the attached sensor head (model, serial, type, wavelength range, power ranges), and report the present wavelength, averaging, range and zero value.
log_power_series👁 readLog a series of power readings to characterise laser stability, warm-up or drift. Returns mean, stdev, min/max, RMS and peak-to-peak stability (%) and drift (%/min), plus the readings (downsampled to max_points; use save_path, a new .csv file, to keep all of them).
read_power👁 readRead the optical power once (in W, plus dBm and a formatted string). The reading uses the configured correction wavelength and averaging; set the wavelength first.
read_sensor_temperature👁 readRead the sensor head temperature in °C (thermal sensors and other heads with a built-in temperature sensor only).
reconnect🛑 safetyClose and re-open the connection to the instrument (e.g. after it was power cycled or a cable was re-plugged).
set_averaging🎛 controlSet how many samples the meter averages for each reading. More averaging lowers noise but slows each reading (PM100 series: ~3 ms per sample; PM400/PM101/PM5020: 1 ms).
set_range🎛 controlSelect auto-ranging, or a fixed power range that fits range_w (the meter picks the most sensitive range that can hold that power). A fixed range reports over-range if exceeded.
set_wavelength🎛 controlSet the correction wavelength (nm) the meter uses to convert sensor signal to power. Must be within the sensor's calibrated range (see get_device_info); sent as a whole number of nm.
zero_sensor🎛 controlDark-zero the sensor (removes dark current / thermal offset). BEFORE calling, ask the user to block the beam and cover the sensor aperture completely, and wait for them to confirm: zeroing with light on the sensor corrupts all later readings. Takes a few seconds.

get_connection_info, get_command_log and reconnect are built into every LabMCP server.

A power meter only measures, so this server has no hazard tools. Your laser is the hazard: wear the right eyewear and never exceed the sensor's rated power or power density.

Safety limits

LimitDefaultMeaning
max_series_duration_s600 sLongest power-logging series an agent may start, estimated as count × max(interval, reading time), where a reading takes averaging × 3 ms plus ~10 ms

Override at launch: --limit max_series_duration_s=3600. One call can log at most 3640 s (the tool's timeout minus a margin): longer estimates are refused, and a series that runs slower than estimated stops early and returns what it measured (completed: false). save_path must end in .csv and must not exist yet (an existing file is never overwritten); it is checked before logging starts.

Example prompts

  • "Set the meter to 1064 nm and tell me the power after the objective."
  • "Log the laser power every 2 s for 10 minutes and report the RMS and peak-to-peak stability. Save the data to ~/data/laser_warmup.csv."
  • "I've capped the sensor. Zero it, then I'll remove the cap and you read the power."
  • "Measure the transmission of this ND filter: read the power without it, then again after I insert it, and give me the optical density."
  • "Is the thermal sensor head still warming up? Read its temperature every minute until it's stable."

Notes

  • Wavelength matters. Photodiode responsivity varies by a factor of several across the sensor's range, so a wrong wavelength setting gives a wrong power. The server sends whole nanometres (the PM400/PM101/PM5020 references specify an integer), so 632.8 nm is set as 633 nm. That changes the responsivity by far less than the sensor's calibration uncertainty.
  • Units. MEAS:POW? answers in the meter's configured unit (W or dBm). The driver reads SENS:POW:UNIT? first and always returns watts, so a front-panel unit change can't mislead the agent.
  • Out of range. The meter answers INFINITY / 9.9E37 when the signal exceeds a manually selected range. The server reports this as an error that suggests auto-ranging.
  • Zeroing (SENS:CORR:COLL:ZERO) runs in the background. The server polls …:ZERO:STAT?, checks the error queue, and returns the zero value (…:ZERO:MAGN?). Zero values are lost when the meter reboots.
  • Energy (pyroelectric) sensors are recognised but not supported: single-pulse energy measurement needs a trigger/fetch workflow that this server doesn't implement yet.
  • Averaging is SENS:AVER:COUN. One sample takes ~3 ms on the PM100 series and 1 ms on the PM400/PM101/PM5020. The reply timeout scales with the averaging count.
  • PM101 and PM5020: Thorlabs' command references for these list exactly the commands used here. The PM5020 adds a channel suffix (SENS2, MEAS2, SYST:SENS2:IDN?), selected with --option channel=2. Neither has been tested yet, and PM5020 LAN/serial connections have not been tried. The PM100D's Windows TLPM DLL driver is not used; this server speaks SCPI only.
  • The simulator's error replies for out-of-range values (-222,"Data out of range") use the standard SCPI code. The Thorlabs documents only list -113,"Undefined header" explicitly.

Hardware verification

ModelFirmwareInterfaceVerified byDate
*none yet: be the first*